• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 MAST ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 MAST ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-88-60
Device Problems Pumping Stopped (1503); Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 02/05/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 mast roller pump.The incident occurred in (b)(6).A serial readout was provided to livanova (b)(4) for further investigation.During the evaluation the reported issue could be confirmed.Therefore livanova (b)(4) requested the pump and the s5 panel back for further evaluation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a s5 mast roller pump stopped shortly after the speed was increased during priming.Further the customer experienced that the pump lost the connection to the s5 panel.There was no patient involvement.
 
Event Description
See initial.
 
Manufacturer Narrative
A review of the dhr's could not identify any deviations or nonconformities relevant to the issue.The reported issue could not be reproduced during investigation at the manufacturer site.Therefore the root cause could not be identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 MAST ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8389416
MDR Text Key137852961
Report Number9611109-2019-00128
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K062396
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 04/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-88-60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/05/2019
Initial Date FDA Received03/05/2019
Supplement Dates Manufacturer Received03/06/2019
Supplement Dates FDA Received04/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-