Model Number 10-88-60 |
Device Problems
Pumping Stopped (1503); Connection Problem (2900)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/05/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the s5 mast roller pump.The incident occurred in (b)(6).A serial readout was provided to livanova (b)(4) for further investigation.During the evaluation the reported issue could be confirmed.Therefore livanova (b)(4) requested the pump and the s5 panel back for further evaluation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a s5 mast roller pump stopped shortly after the speed was increased during priming.Further the customer experienced that the pump lost the connection to the s5 panel.There was no patient involvement.
|
|
Event Description
|
See initial.
|
|
Manufacturer Narrative
|
A review of the dhr's could not identify any deviations or nonconformities relevant to the issue.The reported issue could not be reproduced during investigation at the manufacturer site.Therefore the root cause could not be identified.
|
|