STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PRIM CEM FXD BPLT #5; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
Catalog Number 5520B500 |
Device Problem
Migration (4003)
|
Patient Problem
Injury (2348)
|
Event Date 02/05/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
A review of the device history records indicates that the reported devices were manufactured and accepted into final stock with no reported discrepancies.The complaint history review indicated that there was no other similar events for the reported lot.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
|
|
Event Description
|
It was reported that the patient's right knee was revised due to lateral subsidence of the tibial baseplate.Intra-operatively, the tibial component was noted to be loose.The baseplate and 5x11 insert were revised to a size 5 universal baseplate with a 15x100 cemented stem, two augments, and a 5x13 cs insert.Rep reported that x-rays, medical records, and additional information are not available due to hospital policy.
|
|