BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ NS FILLED SYRINGE; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 306546
Device Problem Flushing Problem (1252)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/19/2019
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Investigation summary: since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident.No root cause can be determined as no samples were received.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Based on no sample, the investigation concluded, bd was not able to verify the indicated failure.A device history record review was completed with zero defects found.No quality notifications were written for this batch, nor for the associated assembly batches.Investigation conclusion: this is the 2nd complaint for the lot# 8197649 for the same defect or symptom.Previous complaint (b)(4).There was no documentation of issues for the complaint of batch 8197649 during this production run.Root cause description: undetermined.Rationale: na.
 
Event Description
It was reported that bd posiflush¿ ns filled syringe stopped pushing during a flush.This occurred on 4 separate occasions.No serious injury or medical intervention was reported.Verbatim: "material no.: 306546 batch no.: 8197649.It wa reported that the pre-filled syringe stopped pushing during a flush.Per complaint form: during the flushing process in pet/ct unit, about 5 bd prefilled 10cc saline syringes were encountered that stopped pushing before the syringe was empty.While pushing in the flush, they were able to get in about 5 or 6 cc¿s, and then it stopped.They weren¿t able to push anymore flush in.It locked up and they weren¿t able to push or pull the plunger in either direction.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ NS FILLED SYRINGE
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652341
MDR Report Key8396074
MDR Text Key138254562
Report Number1911916-2019-00242
Device Sequence Number1
Product Code NGT
UDI-Device Identifier30382903065463
UDI-Public30382903065463
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141311
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/31/2021
Device Catalogue Number306546
Device Lot Number8197649
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/19/2019
Initial Date FDA Received03/06/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/16/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-