Catalog Number 1885 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/12/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device history record of batch number 74f1801134 that belong to catalog number 1885 has been reviewed and no issues or discrepancies were found which could potentially be related to this complaint.The device has been received for evaluation, however the evaluation is still in progress at the time of this report.A follow up report will be submitted at the conclusion of the device investigation.
|
|
Event Description
|
Customer complaint states "head nurse found the cup leakage when the medicine was injected into the cup prior to use on patient.Replaced new one to complete the treatment.Patient is fine.".
|
|
Event Description
|
Customer complaint states "head nurse found the cup leakage when the medicine was injected into the cup prior to use on patient.Replaced new one to complete the treatment.Patient is fine.".
|
|
Manufacturer Narrative
|
(b)(4).The sample was returned for evaluation.A visual inspection was performed and damage on the nebulizer jar was observed.No leakage i ssues could be confirmed as a functional inspection test could not be performed due to damage on the nebulizer jar.The received sample does not produce mist at all due to damage on the nebulizer jar, and there is not sufficient evidence to assure this issue was originated during the manufacturing process.The root cause for the condition reported could not be identified.Teleflex will continue to monitor and trend for reports of this nature.
|
|
Search Alerts/Recalls
|