It was reported that a hahn tapered implant driver malfunctioned.The doctor reported that on (b)(6) 2019, while placing an implant at tooth location #22, the implant driver snapped during final torque.The implant driver would slip and skip off of the implant while torqueing the implant down.The implant could not be completely torqued and placed due to the malfunction of the implant driver.Therefore, the implant had to be removed.The doctor immediately placed a slightly larger implant at that tooth location.The patient experienced minor discomfort.There was no report of injury to the patient, and the patient is feeling better.The doctor reported that after closer inspection of the implant driver under a microscope, he noted that there was a flattened and deformed area where the implant driver connects to the implant.There was no abnormality noticed with the implant itself.The implant driver was manufactured on 06/16/2017, and was purchased by the doctor in (b)(6) 2017.
|
The device was returned and evaluated.A visual and microscopic inspection were performed on the returned device.There was no fracture identified with the device, the driver was intact.It was observed that the surface of the hex was worn out and flatten.The complaint was confirmed, as a defect with a worn or deformed hex will impact torqueing of an implant.There were no other complaints reported for this lot.This event will be tracked and trended.
|