ETHICON ENDO-SURGERY, LLC. 75MM SELECTABLE NEW TLC; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number NTLC75
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/02/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).Batch # r58758.Device evaluation summary: the analysis results found that the ntlc75 device was received with the firing mechanism damaged as the firing knob was detached and the slip block assembly was noted to be damaged.The device was received with one reload present.The reload was returned with the knife exposed and the proximal 38 drivers up without staples, and the remaining drivers down with staples present; the swing tab in the locked position.In addition, the doghouse of the reload was noted to be damaged.No functional test could be performed with it due to the condition of the device.The damage to the firing knob and slip block assembly is consistent with high (outside indicated use) staple forming forces; however, there is insufficient evidence to determine the cause of the higher loads.It should be noted that the cartridge reload is designed to lockout, as a safety feature, if any staples have been fired from the cartridge reload.If enough force is applied the device could be damaged.For additional information please refer to the instructions for use.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Event Description
It was reported that in the course of a procedure (rt hemi-colectomy) using the ntlc , an error occurred in the process of firing.The blade suddenly stopped in the middle of the device, and the blade didn't move forward.As an emergency measure, the surgeon backed off the firing knob and detach it from the tissue.It is unknown what was used to complete the procedure.There were no reported adverse consequences for the patient so far.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
75MM SELECTABLE NEW TLC
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key8403540
MDR Text Key138554000
Report Number3005075853-2019-17071
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10705036002628
UDI-Public10705036002628
Combination Product (y/n)N
PMA/PMN Number
K092577
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 01/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/24/2023
Device Catalogue NumberNTLC75
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2019
Initial Date Manufacturer Received 03/04/2019
Initial Date FDA Received03/08/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/24/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-