Catalog Number 8065977763 |
Device Problems
Crack (1135); Positioning Problem (3009); Scratched Material (3020)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/14/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Evaluation summary: the product was not returned for analysis.The product history records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There has been one other similar complaint reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A facility representative reported that during an intraocular lens (iol) implant procedure, the doctor noticed that the lens loader was cracked and would not load the lens.They then re-inserted the same lens into a second lens loader and realized that the lens was scratched.They then had to use a third loader.There are two medical device reports associated with this report.This report is for the first cartridge used.
|
|
Manufacturer Narrative
|
Corrected information provided.The manufacturer internal reference number is: (b)(4).
|
|