• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS X-TEN; LIGHT, SURGICAL, CEILING MOUNTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS X-TEN; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Device Problems Break (1069); Particulates (1451)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/27/2019
Event Type  malfunction  
Manufacturer Narrative
The issue is being investigated by manufacturing site.(b)(4).
 
Event Description
On (b)(4) 2019 maquet (b)(4) became aware of an issue with one of the surgical light- x-ten.As it was stated, the plastic particles were crumbling from the direction handles of the device.There was no injury reported, however it was decided to report this issue based on the potential as any particles falling off into the sterile field or during operation might be a source of contamination.(b)(4).
 
Manufacturer Narrative
The issue is still being investgated by manufacturing site.
 
Event Description
(b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer reference number: (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer reference number: (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer reference number: (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer reference number: (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer reference number: (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer reference number: (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer's reference number: (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer reference number: (b)(4).
 
Manufacturer Narrative
Getinge became aware of an issue with one of the surgical light- x-ten.As it was stated, the plastic particles were crumbling from the direction handles of the device.There was no injury reported, however it was decided to report this issue based on the potential as any particles falling off into the sterile field or during operation might be a source of contamination.It was established that when the event occurred, the surgical light did not meet its specification and it contributed to the event.The provided information did not indicate that the device was being used for patient treatment when the event took place.During the investigation it was found that there is no apparent trend in the complaints.Despite several reminders it was not possible to get information about the chemicals used for cleaning.Nevertheless, for such a loss of mechanical characteristics, the hypothesis of a chemical root cause is very likely.The age of these handles combined with not compliant chemical product is the most likely root cause of this phenomenon.We believe the related devices are performing correctly in the market.We also believe that if the manufacturer recommendation would have been followed the incident would have been avoided.
 
Event Description
Manufacturer reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-TEN
Type of Device
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
orléans cedex 2
FR 
MDR Report Key8404251
MDR Text Key138538250
Report Number9710055-2019-00063
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
PMA/PMN Number
K040735
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Inspection
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 12/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/27/2019
Initial Date FDA Received03/08/2019
Supplement Dates Manufacturer Received02/27/2019
02/27/2019
02/27/2019
02/27/2019
02/27/2019
02/27/2019
02/27/2019
02/27/2019
09/24/2019
09/24/2019
12/02/2019
Supplement Dates FDA Received04/03/2019
04/26/2019
05/21/2019
06/19/2019
07/12/2019
08/06/2019
08/30/2019
09/24/2019
10/18/2019
11/15/2019
12/06/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-