• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION M SERIES; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION M SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number MSERIES BI-PHASIC
Device Problem Unable to Obtain Readings (1516)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event Description
Complainant alleged that during biomed testing, the device was unable to obtain an ecg signal via electrode pads.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Manufacturer Narrative
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a faulty resistor on the system board.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key8405376
MDR Text Key138290855
Report Number1220908-2019-00565
Device Sequence Number1
Product Code DPS
UDI-Device Identifier00847946002909
UDI-Public00847946002909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K990762
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberMSERIES BI-PHASIC
Device Catalogue NumberM SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2019
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/15/2019
Initial Date FDA Received03/08/2019
Supplement Dates Manufacturer Received02/15/2019
Supplement Dates FDA Received03/20/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-