According to the reporter, they had a study, but it was not on the recorder when they came to download it.It could have been the study was never started.The recorder worked correctly during the previous procedure.There was no patient harm.A repeat procedure on a different day was necessary.
|
Evaluation summary: this report is based on information provided by the complaint tracking system.According to cs text in gch a product sample of bravo recorder was provided by customer to medtronic cs at heerlen site unit but was not provided to yokneam investigation team.Since no sample was returned for investigation it is impossible to determine if product meet specs or not.Since no sample arrived.Visual inspection cannot be performed.The customer reported that they had a study, but it was not on the recorder when they came to download it.It could have been the study was never started.The investigation found per ts: bravo recorder has been tested and 48 hours study was successful.The investigation found the device to function normally and within specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
|