The embolic material will not be returned for analysis as it was consumed in the intervention.Since the device was not returned, we are unable to perform further root cause analysis.All devices are 100% tested and all products are 100% inspected for damages and irregularities during manufacture.Linked with mdr: 2029214-2019-00190.If information is provided in the future, a supplemental report will be issued.
|
Medtronic received report that during the treatment of a distal dural fistula, the marathon catheter encountered resistances when attempted to be retrieved, after the liquid embolization was completed.The catheter was reported to have stretched and separated at the distal segment.The distal segment was unable to be removed and was left within the ophthalmic and internal carotid artery.However, the vessels remained patent.The patient was asymptomatic post the embolization procedure.During the procedure, there was 45 seconds pause between onyx injections and 2cm of onyx reflux was noted at the distal tip of the catheter.The devices were reported to have been prepared and used per the instructions for use (ifu).The vessel was normal tortuosity.
|