Catalog Number 114916
Device Problems Fracture (1260); Loose or Intermittent Connection (1371); Appropriate Term/Code Not Available (3191)
Patient Problems Bone Fracture(s) (1870); No Known Impact Or Consequence To Patient (2692)
Event Date 02/12/2019
Event Type  Injury  
Event Description
Revision surgery - due to the patient fracturing their humerus, making implant loose.The surgeon replaced humeral stem and bearings during revision.
 
Manufacturer Narrative
The reason for this revision surgery was due to loosening.The previous surgery and the surgery detailed in this investigation occurred 1.3 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to loosening.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISCOVERY ELBOW
Type of Device
DISC HUM 5X150MM LT FLANGED C
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8412935
MDR Text Key138571251
Report Number1644408-2019-00171
Device Sequence Number1
Product Code JDC
UDI-Device Identifier00888912225564
UDI-Public(01)00888912225564
Combination Product (y/n)N
PMA/PMN Number
K013042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 05/22/2019
2 Devices were Involved in the Event: 1   2  
2 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number114916
Device Lot Number708300
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/12/2019
Initial Date FDA Received03/12/2019
Supplement Dates Manufacturer Received05/16/2019
Supplement Dates FDA Received05/22/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number0
Patient Outcome(s) Other; Required Intervention;
Patient Age67 YR
-
-