Model Number N/A
Device Problems Fracture (1260); Device Damaged by Another Device (2915)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Concomitant medical products: catalog #: 406669, comprehensive threaded steinmann pin, lot # unknown.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2019-01085.Will be returned.
 
Event Description
It was reported that the pin broke while using short quick-release drill chuck device.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).Reported event was confirmed as the pin was returned fractured inside the drill.And an analysis states that states that the threaded tip sample fractured due to torsional overload.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VNGD SHORTQUICK-RELEASE DRILL
Type of Device
PROSTHESIS KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8413727
MDR Text Key138573768
Report Number0001825034-2019-01084
Device Sequence Number1
Product Code LXH
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number32-486259
Device Lot NumberZB180101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/15/2019
Initial Date FDA Received03/12/2019
Supplement Dates Manufacturer Received08/20/2019
Supplement Dates FDA Received08/28/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-