TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
|
Back to Search Results |
|
Model Number 500AHCT |
Device Problem
Failure to Power Up (1476)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/16/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reported complaint was identified during unit check-up.During laboratory analysis, the product surveillance technician (pst) observed that the monitor would not power on because of defective auxiliary (aux) board.The aux board was found to have an overheated capacitor.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
It was reported that during the use of the device for a non-clinical activity, the blood parameter monitor (bpm) had no power and would not boot up.There was no patient involvement.
|
|