• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 NEUTRAL ARCOMXL LNR 36MM F; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 NEUTRAL ARCOMXL LNR 36MM F; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Difficult or Delayed Positioning (1157)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/12/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Reported event was confirmed by evaluating the returned device.Visual inspection of the liner identified the damages on the locking feature as well as scallops.There were scratches on the rim of the liner.These damages were likely caused during an attempt to seat the liner into the shell and during removal.Dimensional analysis cannot be performed on the liner due to its impaction and resulting damage.Review of the device history record identified no deviations or anomalies would contribute to reported event.A definitive root cause cannot be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the liner did not seat into the shell as required during an unknown surgery.The liner was removed and replaced with a second device for completion.No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 NEUTRAL ARCOMXL LNR 36MM F
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key8416104
MDR Text Key138815974
Report Number0001825034-2019-01180
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial
Report Date 03/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/07/2023
Device Model NumberN/A
Device Catalogue Number010000741
Device Lot Number6394607
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/01/2019
Initial Date FDA Received03/13/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/19/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-