• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number 87035
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2019
Event Type  malfunction  
Manufacturer Narrative
The device was returned for analysis.Visual inspection found separation of the sensor as well as a bent deployment shaft.X-ray showed a break in the magnetic sensor.The magnetic sensor separated at the white/black collar allowing the wires to fracture creating an open sensor.
 
Event Description
Reportable based on device analysis completed on (b)(6) 2019.It was reported that the magnetic sensor was defective.No further details were reported.However, device analysis found that the magnetic sensor was separated at the white/black collar.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
302 parkway
global park, la aurora,
heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330,
saint paul, MN 55112
6515826168
MDR Report Key8419911
MDR Text Key138765461
Report Number2134265-2019-02467
Device Sequence Number1
Product Code DRF
UDI-Device Identifier08714729841968
UDI-Public08714729841968
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K143481
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/19/2019
Device Model Number87035
Device Catalogue Number87035
Device Lot Number0022755300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/28/2019
Initial Date FDA Received03/14/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/05/2018
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-