MAUDE Adverse Event Report: MAKO SURGICAL CORP. 3.0 RIO® ROBOTIC ARM - MICS; STEREOTAXIC DEVICE, ROBOTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. 3.0 RIO® ROBOTIC ARM - MICS; STEREOTAXIC DEVICE, ROBOTICS Back to Search Results
Catalog Number 209999
Device Problems Computer Software Problem (1112); Non Reproducible Results (4029)
Patient Problem Discomfort (2330)
Event Date 03/04/2019
Event Type  Injury  
Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
 
Event Description
This pi is for the robot used in the primary surgery.It was reported that the patient's left hip was revised due to the shell shifting and becoming anteverted.Surgeon confirmed shifting by comparing post-op x-rays from primary implant on (b)(6) 2019 and post-op x-rays in late (b)(6) 2019.Possible cause or contributor to shell shifting was not reported to rep.A shell, 2 screws, liner, biolox ceramic head and accolade ii stem were revised to a trident tritanium revision shell with 5 screws, an adm/mdm liner construct, lfit v40 metal head, and accolade ii stem.There are no allegations against the liner, head, or stem.Rep provided primary and revision implant sheets and reported no further information will be released by the hospital or surgeon.
 
Manufacturer Narrative
Follow-up #1 and final report submitted to update sections d.3, g.1, g.4, g.7, h.2, h.3, h.6, h.10 and h.11 based on the results of investigation.Reported event: an event regarding revision involving 3.0 rio robotic arm - mics, catalog: 209999 was reported.Method & results: -device history review: not performed as the device being inspected is software.Rio serial number not reported.-complaint history: based on the device identification ((b)(4)) the complaint databases were reviewed from 2011 to present for similar reported events regarding revision.There were 8 other reported event for the listed catalog number ((b)(4)).-conclusion: product inspection could not be completed due to no logs or session files being released by the hospital or surgeon.H3 other text : evaluation could not be performed because the logs/session files were not provided.
 
Event Description
This pi is for the robot used in the primary surgery.It was reported that the patient's left hip was revised due to the shell shifting and becoming anteverted.Surgeon confirmed shifting by comparing post-op x-rays from primary implant on (b)(6), 2019 and post-op x-rays in late (b)(6) of 2019.Possible cause or contributor to shell shifting was not reported to rep.A shell, 2 screws, liner, biolox ceramic head and accolade ii stem were revised to a trident tritanium revision shell with 5 screws, an adm/mdm liner construct, lfit v40 metal head, and accolade ii stem.There are no allegations against the liner, head, or stem.Rep provided primary and revision implant sheets and reported no further information will be released by the hospital or surgeon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.0 RIO® ROBOTIC ARM - MICS
Type of Device
STEREOTAXIC DEVICE, ROBOTICS
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key8425998
MDR Text Key138970936
Report Number3005985723-2019-00237
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00848486030407
UDI-Public00848486030407
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 07/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number209999
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/04/2019
Initial Date FDA Received03/15/2019
Supplement Dates Manufacturer Received06/30/2019
Supplement Dates FDA Received07/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age60 YR
-
-