Model Number 60-01-00
Device Problem Decoupling (1145)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/14/2019
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in(b)(6).If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
 
Event Description
Livanova (b)(4) received a report that a centrifugal pump system with tubing clamp decoupled independently according to the perfusionist, with impact on flow and speed settings during procedure.
 
Manufacturer Narrative
H10: visual inspection revealed that the locking pin of the centrifugal pump system with tubing clamp was not right in place and that the main cable was not fixed well.During detailed investigation, the described error could not be reproduced.As the issue could not be reproduced, no specific root cause could be determined.Although, the identified deviation in the locking mechanism might have contributed to the reported malfunction.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8428257
MDR Text Key139323385
Report Number9611109-2019-00163
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-01-00
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/18/2019
Initial Date FDA Received03/18/2019
Supplement Dates Manufacturer Received06/12/2019
Supplement Dates FDA Received07/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-