Model Number NR440Z
Device Problems Product Quality Problem (1506); Defective Device (2588); Device Contaminated During Manufacture or Shipping (2969)
Patient Problem Foreign Body Reaction (1868)
Event Date 02/21/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Implant date: unknown.When additional information is received a follow up report will be submitted.
 
Event Description
It was reported by the healthcare professional "patient had an enduro implanted, presented with joint pain and stiffness, was examined in clinic and determine an exploratory washout was necessary.Intraoperative tissue samples were collected and sent for lab work - tissues had a blackened appearance.The underside of the patellar tendon and intercondylar box had the highest concentration of blackened tissue.No damage to any of the components was observed neither were there indications of loosening of the hinge mechanism.A thorough washout was performed and the wound closed.Surgeon awaiting lab results to determine treatment.Images from case are available." this incident did not cause or contribute to serious injury or death or a delay in surgery.All medwatch submissions related are: 9610612-2019-00179.
 
Event Description
Clarification was received: it was reported that a patient was initially implanted with enduro knee products in the right knee on (b)(6) 2018.On (b)(6) 2019, the patient underwent an arthrotomy of the knee and synovectomy; antibiotic beads were inserted for treatment of septic right knee.Tissue specimens were taken.Results showed acute inflammation/neutrophilic infiltration, and granulation tissue formation; no definitive evidence of malignancy.No additional patient harm or intervention has been reported since the surgery.Associated components (also reported separately): as femur stem neutral, as enduro femur spacer, as nut for femur stem, as enduro femoral component, as enduro femur spacer, patella, as tibia offset stem, as enduro meniscal component, as enduro tibial hemi-wedge x 4, as enduro tibial component.
 
Manufacturer Narrative
Additional information: patient information, event description, relevant tests/laboratory data: tests, common device name, model #, lot #, expiration date, implant date, manufacture date, manufacturing site evaluation: devices are not available because the components are still implanted.No pictures were provided; surgical specimen and final results provided.We do not know the consistency and origin of the mentioned tissue with blackened appearance.Batch history review: the device quality and manufacturing history records have been checked for all available lot numbers and found to be according to our specifications valid at the time of production.Conclusion and root cause: based on the information available, it is not possible to determine a possible root cause for the failure.At this time we assume that the failure is not product related.It could be possible that the failure is usage/patient related.Rationale: in light of the incomplete information received about the lab results, and due to the circumstance that we did not receive the complained devices for investigation, it is not possible to determine a root cause for the mentioned failure.There are no hints for a material problem.According to the quality standard and files a material defect and production error is improbable.At this time we exclude a design-related error because the market feedback is unremarkable on this matter.A maintenance failure also can be excluded because this issue is only relevant for instruments.It could be possible that due to so-called "third-body wear" (such wear includes foreign particles for example bone cement, bone, or metal particles) in the knee mechanism (bush bearing) lead to a peek abrasion which could lead potentially to the mentioned blackened tissue.A capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AS FEMUR EXTENS.STEM 5° D20X177 CEM.LESS
Type of Device
KNEE ENDOPROSTHESES
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key8434527
MDR Text Key139290458
Report Number9610612-2019-00178
Device Sequence Number1
Product Code KRO
Combination Product (y/n)N
PMA/PMN Number
K101815
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 06/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNR440Z
Device Catalogue NumberNR440Z
Device Lot Number52171650
Was Device Available for Evaluation? No
Distributor Facility Aware Date04/15/2019
Initial Date Manufacturer Received 02/21/2019
Initial Date FDA Received03/19/2019
Supplement Dates Manufacturer Received04/15/2019
Supplement Dates FDA Received06/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age50 YR
-
-