HOLOGIC, INC. FLUENT FLUID MANAGEMENT SYSTEM; HYSTEROSCOPIC INSUFFLATOR Back to Search Results
Model Number FLT-112
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/01/2019
Event Type  malfunction  
Manufacturer Narrative
Serial number of the disposable device not provided by the complainant.The device is not being returned therefore, a failure analysis of the complaint device cannot be completed.Lot number of the disposable device not provided by the complainant, therefore the manufacture date is not known.Device history record (dhr) review and sterile lot review were unable to be conducted for the disposable device as the lot number was not provided by the complainant.Internal complaint reference: (b)(4).
 
Event Description
It was reported that during a 3cm fibroid removal using the fluent fluid management system there was initially difficulty distending the cavity and the cervical seal became compromised.Fluid was leaking from the cervix and filling the under buttock drape causing the fluid deficit to rise.The physician removed the hysteroscope and the pump was stopped while he attempted to gain access to the cervix again.The scope was reinserted and a second tenaculum was used to seal the cervix.Distention improved and tissue resection began.About 80-90% of the fibroid was removed when a loud pop was heard and it was noted that the "tissue sock exploded and fluid and specimen fell to the floor." the tissue sock was exchanged as well as the waste bag.The procedure was terminated at a fluid deficit of 1467ml.A telfa was used to retrieve the remaining specimen that was on the floor.No injury or misdiagnosis was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENT FLUID MANAGEMENT SYSTEM
Type of Device
HYSTEROSCOPIC INSUFFLATOR
Manufacturer (Section D)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
kristin fornieri
36 & 37 apple ridge road
danbury, CT 06810
2037318491
MDR Report Key8441287
MDR Text Key140004125
Report Number1222780-2019-00060
Device Sequence Number1
Product Code HIG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K180825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 03/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFLT-112
Device Catalogue NumberFLT-112
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/01/2019
Initial Date FDA Received03/21/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-