Investigation: based on the current available information it is unknown at this time if air is returned to the donor during the pressure test.The reported alarm occurs during ac prime, so the donor should not have been connected at this point in the procedure.In this event, the donor was connect prematurely (indicated by blood in the diversion pouch) and this gives a potential of pushing air back to the donor during the pressure test.The customer provided the photographs of the used trima disposable set to terumo bct for evaluation.Upon photographic inspection, it was confirmed that there was an alarm that occurred as specified by the customer.The reported alarm indicates for ac line obstruction failure-ac occlusion during inlet prime.This alarm checks for ac pump errors and ac occlusion errors and will discontinue the procedure.Also, through pictures it was confirmed that blood and air were interspersed throughout the inlet coil, cassette and return reservoir.The sample line pinch clamp was observed to be closed.The sample bag was full and air was present in the bag.The ac bag frangible appeared to be fully broken and ac was observed within the drip chamber and throughout the ac pump header and inlet coil.A review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.Investigation is in process.A follow-up report will be provided.
|
This report is being filed to provide in investigation: the run data file (rdf) was anaylzed for this event.Root cause: based on the rdf analysis, this alarm #10 occurs during ac priming, where there isno risk for air to be returned to the donor if the donor was connected early.During ac prime,the pumps only run to draw or pull fluid into the cassette, nothing is pushed toward the needleline.The alarm #10, ac line obstruction failure, criteria checks for a decrease in pressure tosignify if the ac line cannot successfully pull fluid.In this case, the pressure signals increased whenthe donor was connected at the prompt to connect ac, then decreased with the pumpmovement during ac prime.The decrease in the pressure signals matched the signals as if ac fluidflow was not satisfactory.The alarm forces the operator to terminate the procedure, socontinuing to the first return is also not possible.Based on the information provided, rdf analysis, the evaluation of the returned disposable, andthe reported alarm this occurrence was due to premature connection of the donor.This isconsidered a user interface error.
|