The customer reported that during a patient procedure, using a glidescope gvl single use size 4 stat, the tip broke off in the patient's mouth.The stat and the broken portion of the stat were removed from the patient.The customer indicated the procedure was completed; however, type of device used to complete the procedure and the length of time to prepare the second device were not reported.No harm to the patient or user was reported.
|
The customer returned two samples of their glidescope gvl single use size 4 stats to verathon for evaluation.A verathon technical service representative evaluated the returned stats where they were able to verify the reported damage.It was found that the ends of the stats had broken off inside the packaging.The returned stats were forwarded to the contract manufacturer for further evaluation.The contract manufacturer evaluated the returned stats where they confirmed the damage to the unit.A visual inspection of the stat was performed and no obvious signs of stress on the stat were observed.The pouch showed minor signs of stress, including a small puncture, indicating potential damage during shipment; however, there were no obvious blow outs or tears in the pouches.The contract manufacturer load tested all available retained samples from this lot and all stats passed inspection.The retain samples were re-measured to inspect for any possible thin sections of the stat.There were no out-of-specification or outlying measurements noted.The machine settings for all retain molding runs were reviewed and all settings were within specification.The sterilization process was also reviewed, and all applied minimum and maximum doses were normal.The contract manufacturer and verathon were unable to definitively determine a root cause for the breaks observed.No corrective action is required at this time.Verathon will continue to monitor for trends.
|