• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/27/2019
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed as per company standard operating procedure since the device manufacture date is greater than one year from the event date.A getinge service territory manager (stm) was dispatched to investigate.The stm confirmed the iabp unit had no ecg signal when using the customer's ecg cables during the evaluation.The stm performed a full performance check with known working ecg cables, and all tests passed and the stm did not find any issues with the iabp unit.The stm disposed of the non working ecg cabled and ordered a spare set of ecg cables.Subsequently, the stm completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was released to the customer and cleared for clinical use.The first name of the initial reporter has been abbreviated due to field character limit; the full name should read (b)(6).The full name of the event site was shortened due to field character limit; the full name is (b)(6).
 
Event Description
It was reported that during use on a patient, the ecg cable on the cardiosave intra-aortic balloon pump (iabp) was not registering a waveform.There was no patient harm and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key8445457
MDR Text Key139712678
Report Number2249723-2019-00460
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/28/2019
Initial Date FDA Received03/22/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-