The lot number was provided.A review of the approved device history records indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The device has not been returned to the manufacturer; therefore, a product analysis could not be performed.The device was used in the same procedure as was reported on mfr.Report number 2032493-2019-00063.The instructions for use (ifu) identifies premature or difficult coil detachment as potential complications associated with use of the device.
|