• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 PINN STRAIGHT CUP IMPACTOR; HIP INSTRUMENTS : IMPACTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 PINN STRAIGHT CUP IMPACTOR; HIP INSTRUMENTS : IMPACTORS Back to Search Results
Catalog Number 221750041
Device Problem Break (1069)
Patient Problems No Consequences Or Impact To Patient (2199); Foreign Body In Patient (2687)
Event Date 03/04/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the bottom ring of impactor was broken off.No surgical delay.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN STRAIGHT CUP IMPACTOR
Type of Device
HIP INSTRUMENTS : IMPACTORS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8445647
MDR Text Key139681677
Report Number1818910-2019-88254
Device Sequence Number1
Product Code MRA
UDI-Device Identifier10603295098980
UDI-Public10603295098980
Combination Product (y/n)N
PMA/PMN Number
P070026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 03/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number221750041
Device Lot NumberSO2029660
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2019
Initial Date Manufacturer Received 03/04/2019
Initial Date FDA Received03/22/2019
Supplement Dates Manufacturer Received04/03/2019
05/01/2019
Supplement Dates FDA Received04/15/2019
05/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age65 YR
-
-