SMITH & NEPHEW, INC. FEMORAL IMPLANT IMPACTOR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
Catalog Number 71440005 |
Device Problem
Flaked (1246)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/27/2019 |
Event Type
malfunction
|
Event Description
|
It was reported that the black coating came off.No delay or injury reported.A backup device was available.
|
|
Manufacturer Narrative
|
The associated devices were returned and evaluated.A visual inspection of the returned devices confirmed the stated failure mode.The coating was missing on the anterior arms and was peeling off of the posterior arms.The devices were manufactured in 2007 and 2009 and exhibit signs of extensive wear/ usage.This failure mode has been previously identified.Since the manufacturing of the complaint device, a design change was implemented to eliminate/reduce the occurrence of this failure.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
|
|