• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS CENTRIFUGAL SYSTEM; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS CENTRIFUGAL SYSTEM; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM Back to Search Results
Model Number 6379
Device Problem Unable to Obtain Readings (1516)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/05/2019
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) replaced the flow printer circuit board (pcb).The unit operated to the manufacturer's specifications.The suspect part was returned to the manufacturer for further evaluation.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the unit was not reading any flows.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
Updated event problem and evaluation codes.The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed the module displayed flow and signal strength of over 100% with the flow printed circuit (pc) board installed in a lab-use centrifugal control module.No loss of flow display occurred during the two days of testing.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS CENTRIFUGAL SYSTEM
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key8448982
MDR Text Key139738511
Report Number1828100-2019-00151
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
PMA/PMN Number
K950739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6379
Device Catalogue Number6379
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/05/2019
Initial Date FDA Received03/25/2019
Supplement Dates Manufacturer Received05/09/2019
Supplement Dates FDA Received05/22/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-