MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC SCALPEL HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC SCALPEL HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HP054
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/08/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # r9218w.Investigation summary: the handpiece was received with the nose cone cracked and the mount was noted to be loose.No functional testing could be performed due to the nose cone being extremely cracked and no instrument could be attached to the handpiece.The instrument was disassembled to inspect internal components and the moisture indicator was positive.The transducer assembly was not held in place due to the cracked nose cone, so torquing on the disposable resulted in twisting only the handpiece transducer assembly until the internal wires got disconnected.Due to the cracked nose cone, moisture entered the hand piece mid housing.Analysis was unable to determine the exact root cause that lead to crack in the handpiece nosecone.It is possible that the condition of the nose cone contributed to the assembly issue when trying to attach the instrument to the handpiece.It is possible that the ingress of moisture affected handpiece functionality.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Event Description
It was reported that before a laparoscopic myomectomy procedure, the hp054 was not able to be firmly attached.The nose cone of the handpiece rotated while it was connected to the blade.The number of remaining uses of hp054 was unknown.Handpiece was used to complete the case.There were no adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC SCALPEL HAND PIECE
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key8449841
MDR Text Key139813055
Report Number3005075853-2019-17535
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10705036002147
UDI-Public10705036002147
Combination Product (y/n)N
PMA/PMN Number
K002906
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 03/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHP054
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/13/2019
Initial Date FDA Received03/25/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/17/2018
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-