Model Number TABLETOP-JAPAN |
Device Problem
Computer Software Problem (1112)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/07/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A surgeon reported the system changed from ready to standby during a laser procedure.The console was changed back to ready to complete the procedure.There was no patient harm.
|
|
Manufacturer Narrative
|
The system was examined and the reported event was confirmed.The footswitch was subsequently replaced to resolve the issue.The system was tested and found to meet product specifications.The manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to nonconforming footswitch.However, as to how or when it became non-conforming remains inconclusive.The manufacturer internal reference number is: (b)(4).
|
|