The customer's blood collection tube handling is in line with sarstedt recommendations assuming the radius of their centrifuge is at least 200 millimeters.The clinical sensitivity and specificity section of the advia centaur xp anti-hbs2 instructions for use (ifu) (10629819 revision k) lists the 95% confidence interval (ci) for resolved relative specificity as 98.34% - 99.88%.Given the customer has only had an issue with one sample there is no indication that the advia centaur xp ahbs2 lot 102 is failing to meet the specificity claim in the ifu.A review of internal data for advia centaur xp anti-hbs2 lot 102 indicates it is performing as intended.The cause of the positive result with the advia centaur xp anti-hbs2 lot 102 could not be determined.Siemens cannot rule out normal assay performance, pre-analytical factors, or a sample issue.Based on the investigation, no product problem was identified.The instrument is performing within specifications.No further evaluation of the device is required.
|