|
Model Number 5700006 |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/12/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The referenced device was not returned to olympus for evaluation.The exact cause of the reported event cannot be confirmed.If the device is returned at a later date, this report will be supplemented accordingly.A review of the dhr for this product has been conducted and all records indicate that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.
|
|
Event Description
|
Olympus was informed that during the middle to the end of a therapeutic cystoscopy with removal of a right ureteral stent, the patient had a large, hard stone that required fragmentation for removal.The basket broke when it was removed from the patient.There were no device fragments that broke off or fell into the patient.There was no bleeding reported.A second device from the same lot was used and also broke.The procedure was completed using a third device from a different lot.There was patient injury reported.2 of 2 devices.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to make a correction on the aware date.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide the device evaluation results.The referenced device was returned in the original packaging and the lot number was confirmed as mk715070.A visual inspection was performed on the received condition found the core wire within the handle was kinked.The basket could not be deployed.Additionally the distal end was kinked just below the basket.This suggests that excessive force was applied to the device with the basket retracted.The threaded luer was unscrewed and the sheath could not be removed.The device was disassembled further, removing the wire basket and sheath assembly from the handle.The kink in the core was confirmed and the blue elastic strain relief was cut away from the assembly exposing a buckled sheath.The wrinkled sheath combined with the kinked distal end prevented the device from deploying.After disassemble the basket was manually pulled from inside the sheath.The basket was intact and free from damage.Based on the evaluation, the cause of the kinks in the proximal core and distal end are indicative of misuse.The excessive force applied to the device buckled the proximal end of the sheath preventing operation as designed.This likely occurred during insertion as the device become lodged on anatomy or caught on the scope or urethral stent.The instruction manual for device provides cautions that states, "the surlok helical stone basket is a delicate precision instrument and must be handled with care".
|
|
Search Alerts/Recalls
|
|
|