The customer reported an issue with horizontal table motion; motion halted though the endless screw continued to turn with horizontal backwards motion.The system was removed from service.The philips field service engineer (fse) went to the customer site to evaluate the system.The fse confirmed the reported problem.The customer noticed the couch wouldn¿t move when they attempted to move the table inside or outside across the bore using the gantry control panel or the tape switch.It didn¿t reach the desired position.The system detected the problem and an error appeared.The fse confirmed this issue occurred prior to the start of a scan.Due to this issue, the customer had chosen to stop using the system and called philips service.The philips fse determined the bearings were stuck and didn´t spin correctly.This issue was caused by ball bearing mechanical wear.The fse replaced the ball bearing and returned the system for clinical use.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|