RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
|
Back to Search Results |
|
Device Problem
Protective Measures Problem (3015)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2019.Date of report: 27march2019.No phone number provided.A follow-up report will be submitted once the investigation has been complete.
|
|
Event Description
|
Customer contacted technical support (ts) stating that unit had error message of blower failure the customer reported there was no patient involvement.Event date not specified, estimate used.
|
|
Manufacturer Narrative
|
Date rec'd by mfr: 19nov2019.Date of report: 21nov2019.Failure analysis on the returned bipap focus blower shows that this bipap focus blower assembly was tested and no functional failures were identified.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Manufacturer Narrative
|
G4: 20may2019, b4: 11oct2019.The service engineer replaced the blower assembly to address the reported issue.The unit passed all testing required.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Event Description
|
Customer contacted technical support (ts) stating that unit had error message of blower failure.The customer reported there was no patient involvement.Event date not specified, estimate used.
|
|