• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC (LITTLETON) BASE OARM BI70000028120 SYS 120V; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC (LITTLETON) BASE OARM BI70000028120 SYS 120V; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Model Number BI70000028120
Device Problems Failure to Charge (1085); Overheating of Device (1437); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Patient Involvement (2645)
Event Date 03/04/2019
Event Type  malfunction  
Manufacturer Narrative
The ups power supply was returned to the manufacturer and is currently under analysis.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding an imaging device being used outside a procedure.It was reported that the mobile view station (mvs) was showing a charging error and the power cable to the system was hot to the touch.The biomed was requesting a mvs uninterruptible power supply (ups).There was no patient present when this issue was identified.
 
Manufacturer Narrative
The ups was returned to the manufacturer for analysis.Analysis found that the reported issue was confirmed.The ups failed bench test.The battery was charged for at least 6 hours.The ups failed load test with 2 minutes and 30 seconds.A minimum of 5 minutes is passing.The battery was at the end of its life span.Analysis found that the reported event was related to an electrical issue.The mobile view station (mvs) power cord was discarded by the site.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BASE OARM BI70000028120 SYS 120V
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
MDR Report Key8468106
MDR Text Key140507939
Report Number3004785967-2019-00535
Device Sequence Number1
Product Code OXO
UDI-Device Identifier00643169353428
UDI-Public00643169353428
Combination Product (y/n)N
PMA/PMN Number
K092564
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 04/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI70000028120
Device Catalogue NumberBI70000028120
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2019
Initial Date Manufacturer Received 03/04/2019
Initial Date FDA Received03/31/2019
Supplement Dates Manufacturer Received04/16/2019
Supplement Dates FDA Received04/29/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-