• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 UNKNOWN HIP ACETABULAR CUP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 UNKNOWN HIP ACETABULAR CUP Back to Search Results
Catalog Number UNK HIP ACETABULAR CUP
Device Problems Fracture (1260); Material Separation (1562)
Patient Problems Tissue Damage (2104); No Code Available (3191)
Event Date 09/30/2011
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.This report is for an unknown device/unknown lot.Part and lot number are unknown; udi number is unknown.(b)(4).
 
Event Description
After review of medical records, patient was revised to address failed right hip replacement.Revision notes reported a dislodge polyethylene liner.A broken poly liner as well as a broken well-fixed shell and a well-fixed full coat stem in appropriate version and frozen section was negative for acute inflammation.It was noted that the acetabular shell itself was broken on the anterior aspect of it.A significant amunt of metallosis was also noted.Doi: (b)(6) 2009 - dor: (b)(6) 2011, (right hip).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report. .
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot number was not provided.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP ACETABULAR CUP
Type of Device
HIP ACETABULAR CUP
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8473548
MDR Text Key140626191
Report Number1818910-2019-89325
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
PMA/PMN Number
UNK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup,Followup
Report Date 03/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP ACETABULAR CUP
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/12/2019
Initial Date FDA Received04/02/2019
Supplement Dates Manufacturer Received04/15/2019
05/08/2019
Supplement Dates FDA Received04/17/2019
05/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age62 YR
-
-