• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THERAKOS, INC. CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THERAKOS, INC. CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Model Number NOT APPLICABLE
Device Problems Crack (1135); Leak/Splash (1354); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/20/2019
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction photoactivation module leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot g283 was conducted.There were no non-conformance's associated with this lot.This lot met all release requirements.A review of kit lot g283 for the reported issue shows no trends.Trends were reviewed for complaint categories, photoactivation module leak and alarm #8: blood leak? (photoactivation chamber).No trends were detected for each complaint category.The assessment is based on the information available at the time of investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.Investigation complete.(b)(4).
 
Event Description
The customer called to report a photoactivation module leak during the treatment procedure.The customer stated that approximately 800 ml of whole blood was processed when the leak was observed.The customer stated they received an alarm #8: blood leak? (photoactivation chamber) alarm during the photoactivation phase of the treatment.The customer visually inspected the photoactivation plate and observed it was cracked and leaking.The customer aborted the procedure and did not return blood to the patient.The customer stated the patient was stable.The customer did not return product for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS, INC.
bedminster NJ 07921
Manufacturer (Section G)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
college business & tech park
cruiserath road
blanchardstown, dublin D15 T X2V
EI   D15 TX2V
Manufacturer Contact
megan vernak
1425 us route 206
bedminster, NJ 07921
MDR Report Key8475932
MDR Text Key141189997
Report Number2523595-2019-00047
Device Sequence Number1
Product Code LNR
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 04/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/01/2020
Device Model NumberNOT APPLICABLE
Device Catalogue NumberCLXECP
Device Lot NumberG283
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/20/2019
Initial Date FDA Received04/02/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/26/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-