Model Number 3.0X130MM FIBULA NAIL, LEFT |
Device Problems
Break (1069); Difficult to Remove (1528); Detachment of Device or Device Component (2907)
|
Patient Problem
Device Embedded In Tissue or Plaque (3165)
|
Event Date 03/08/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
The contribution of the device to the reported event could not be determined as the device was requested/is expected but has not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
|
|
Event Description
|
It was reported during a left ankle fracture procedure, the proximal end of the fib30130l fibula nail broke off during insertion and was left inside of the fibula.The rep stated the fracture was successfully reduced and fixated with a distal fibula plate and screws.The seated depth of the implant was 130mm.A 3.22mm was used to prep the bone.Additional information has been requested.Additional information received on (b)(6) 2019: the rep stated that the surgeon attempted to remove the broken fragment, but no avail.The rep confirmed the remaining portion of the nail was retrieved and will be returning for evaluation.The bone was prepped by using a 6.2 distal reamer and 3.2 proximal reamer.The rep stated that the surgeon had to switch to their back up plan to make a larger incision so they could plate the fracture.A second fib30130l was not brought in.The case was completed by using an ankle fracture plate.
|
|
Manufacturer Narrative
|
Complaint confirmed.The nose cone was found to have broken off from the device.The most likely cause is improper canal preparation, which caused the distal end of the device to become incarcerated into the unreamed canal and break when the implant was impacted.
|
|
Search Alerts/Recalls
|