• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH MAQUET HCH 40; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH MAQUET HCH 40; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HEATER-COOLER UNIT
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/26/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.(b)(4).Field service technician (fst) was onsite.It was not possible to reproduce the reported issue.No device-error could be detected.According to the fst the most probable root cause was that the second patient water-connection was short-circuited with a shand, thereby the flow was strongly reduced.Furthermore the fst performed functioning test and a test run.All tests passed.The getinge director of clinical risk management & documentation reviewed the reported issue.In these circumstances, sufficient heat flow could not be ensured.Therefore no product related malfunction can be assumed.Thus the failure could not be confirmed.The issue was caused by an user error.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.
 
Event Description
After active cooling of the patient to 32 degrees celsius, the reheating to 36 degrees celsius over several hours was not possible, so the extra-corporal circulation finally ended at 34 degrees celsius.Device was replaced.No clinical consequences for the patient.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAQUET HCH 40
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key8479889
MDR Text Key141056706
Report Number8010762-2019-00090
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K130300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHEATER-COOLER UNIT
Device Catalogue Number701044054
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/14/2019
Initial Date FDA Received04/03/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-