Siemens has completed an investigation of the reported event.Based on log file analysis and information provided from the customer, the sensis system had no impact on the patient outcome.During the procedure, the real time computer (rtc) displayed all vital signs along with patient waveforms at the monitor.The logs show that several bradycardia and tachycardia were detected multiple times by the system and correctly reported to the users.During the event some non-invasive blood pressure (nipb) measurements, oxygen saturation (spo2) and heart rate (hr) were not logged to file because of a software bug.The failure not to log the vital signs during the event does not result in any patient harm.The vital signs remain visible on the rtc screen throughout the procedure.In addition, logging of the vital signs data will resume when the nibp disturbance (like a kinked measure hose or a loose cuff) is resolved.No further actions are to be taken as there is no negative awareness in regards to the quality and performance of the affected system.
|