• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. PLATINUM 1 SERIES; SURGICAL ADJUNCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. PLATINUM 1 SERIES; SURGICAL ADJUNCTS Back to Search Results
Model Number 1MTEC30
Device Problem Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Date 03/05/2019
Event Type  malfunction  
Manufacturer Narrative
If implanted; give date: n/a (not applicable).The cartridge is not an implantable device.If explanted; give date: n/a (not applicable).The cartridge is not an implantable device; therefore, not explanted.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that the platinum cartridge 1mtec30, lot cd08721, appeared deformed during product handling and prior to insertion.No further information provided.
 
Manufacturer Narrative
Device evaluation: the product was not returned to the manufacturing site; therefore, the complaint issue reported was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Labeling review: the directions for use (dfu) were reviewed.The dfu adequately provides instructions, precautions, and warnings for the proper use and handling of the product.Historical data analysis: a search of complaints revealed 3 other complaints have been received for this production order number.One investigate request is a duplicate which was inadvertently initiated.The other 2 investigations were completed and product deficiency was not confirmed.In addition, there is no increase in trending for the reported complaint type ¿tip deformed¿ for the 1mtec30 models.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATINUM 1 SERIES
Type of Device
SURGICAL ADJUNCTS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key8485665
MDR Text Key141015432
Report Number2648035-2019-00426
Device Sequence Number1
Product Code KYB
UDI-Device Identifier05050474540323
UDI-Public(01)05050474540323(17)190807(10)CD08721
Combination Product (y/n)Y
PMA/PMN Number
K081545
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup,Followup
Report Date 01/01/2005,11/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/07/2019
Device Model Number1MTEC30
Device Catalogue Number1MTEC30
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 03/05/2019
Initial Date FDA Received04/04/2019
Supplement Dates Manufacturer Received05/28/2019
10/25/2020
Supplement Dates FDA Received06/19/2019
11/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-