The patient was undergoing a thrombectomy procedure using an indigo system aspiration catheter 8 kit (kit).During the procedure, the indigo system aspiration catheter 8 (cat8) became fractured near the hub while in use in a non-penumbra sheath.Consequently, the aspiration level decreased, and the cat8 was therefore removed.The procedure was then completed using a new cat8 and the same sheath.There was no report of an adverse effect to the patient.
|
Please note that the device associated with this complaint was expected to be returned; however, additional information received from the penumbra medical affairs associate indicated that the device will not be returned for evaluation.No further information was provided.Therefore, without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|