DEPUY ORTHOPAEDICS, INC. 1818910 SUMMIT SZ6/7 TAPERED REAMER; HIP INSTRUMENTS : REAMERS Back to Search Results
Catalog Number 257002300
Device Problems Difficult to Insert (1316); Device-Device Incompatibility (2919); Material Deformation (2976); Naturally Worn (2988); Positioning Problem (3009)
Patient Problem Not Applicable (3189)
Event Date 03/14/2019
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device is an instrument and is not implanted/explanted.
 
Event Description
It was reported that the end of straight 6/7 summit reamer (d257002300) that mates with hudson adapter would not attach/lock to chuck on t-handle.Post-op assessment of reamer exhibited apparent wear end of reamer that presents reamer from seating clearly in chuck.There was a minute delay.
 
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT SZ6/7 TAPERED REAMER
Type of Device
HIP INSTRUMENTS : REAMERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8487936
MDR Text Key141075057
Report Number1818910-2019-89965
Device Sequence Number1
Product Code HTO
UDI-Device Identifier10603295143024
UDI-Public10603295143024
Combination Product (y/n)N
PMA/PMN Number
UNK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 03/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number257002300
Device Lot NumberSO2039468
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2019
Initial Date Manufacturer Received 03/14/2019
Initial Date FDA Received04/05/2019
Supplement Dates Manufacturer Received04/15/2019
05/09/2019
Supplement Dates FDA Received04/19/2019
05/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-