U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 08april2019.No phone number provided.Manufacturer's field service engineer (fse) evaluated the unit and confirmed the reported issue.The fse replaced the unit's user interface (ui) pcba and battery to resolve the reported issue.Fse performed system calibration and performance verification and all tests passed.Unit ready for service.
 
Event Description
Customer contacted technical support (ts) stating that unit had error code message of power on self test (post) board over temp.Issue was found on post.No patient involvement was reported by the customer.Event date not specified; estimate used.
 
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 25jun19.This complaint is a duplicate of (b)(4).This event was reported on mfr.Report #: 2031642-2019-02160.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM
Type of Device
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key8494021
MDR Text Key141320045
Report Number2031642-2019-02138
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
PMA/PMN Number
K053168
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 03/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/18/2019
Initial Date FDA Received04/08/2019
Supplement Dates Manufacturer Received03/18/2019
Supplement Dates FDA Received06/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-