• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMP RVS TRAY CO 44MM; PROSTHESIS, SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMP RVS TRAY CO 44MM; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Device Problem Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/04/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Device product code: phx.(b)(4).Concomitant medical products: xl-115364, bearing, lot 279990; 115310, glenosphere, lot 115310.
 
Event Description
It was reported that during a primary tsa, the locking ring of the humeral tray would not function correctly.The locking ring was stuck open.The surgeon completed the procedure with another device, with no impact to the patient.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned product/provided pictures identified the bearing is assembled to the tray.The locking ring is stuck open.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.Conclusion statement: if any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP RVS TRAY CO 44MM
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8505758
MDR Text Key142052513
Report Number0001825034-2019-01650
Device Sequence Number1
Product Code HSD
Combination Product (y/n)N
PMA/PMN Number
K113069
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 07/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number115370
Device Lot Number885000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/04/2019
Initial Date FDA Received04/11/2019
Supplement Dates Manufacturer Received06/24/2019
Supplement Dates FDA Received07/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-