Model Number N/A |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/04/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Device product code: phx.(b)(4).Concomitant medical products: xl-115364, bearing, lot 279990; 115310, glenosphere, lot 115310.
|
|
Event Description
|
It was reported that during a primary tsa, the locking ring of the humeral tray would not function correctly.The locking ring was stuck open.The surgeon completed the procedure with another device, with no impact to the patient.Attempts have been made and no further information has been provided.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned product/provided pictures identified the bearing is assembled to the tray.The locking ring is stuck open.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.Conclusion statement: if any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|