Model Number 520-01-008
Device Problem Patient-Device Incompatibility (2682)
Patient Problem Pain (1994)
Event Date 03/14/2019
Event Type  Injury  
Manufacturer Narrative
The reason for this revision surgery was due to pain.The previous surgery and the surgery detailed in this investigation occurred 5.2 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to pain.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event Description
Revision surgery - due to the patient having pain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Type of Device
TURON PRIMARY HUMERAL STEM, SZ 8
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, TX 78758-5445
MDR Report Key8506323
MDR Text Key141673742
Report Number1644408-2019-00304
Device Sequence Number1
Product Code KWS
UDI-Device Identifier00888912144902
UDI-Public(01)00888912144902
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/28/2019
Device Model Number520-01-008
Device Catalogue Number520-01-008
Device Lot Number454G1051
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/14/2019
Initial Date FDA Received04/11/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/02/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
520-00-000, LOT 878C1266; 520-42-116, LOT 913C1026
Patient Outcome(s) Required Intervention;
Patient Age76 YR
-
-