Model Number M00565040 |
Device Problems
Difficult or Delayed Positioning (1157); Premature Activation (1484)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/20/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
Age or date of birth: patient's exact age is unknown; however it was reported that the patient was over the age of 18.(b)(4).The complainant indicated that the suspect device remains implanted and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental mdr will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation on (b)(6) 2019 that a wallflex colonic stent was implanted in the sigmoid colon to treat a 6-7 cm mass during a colonoscopy with stent placement procedure performed on (b)(6) 2019.Reportedly, the patient's anatomy was tortuous and was not dilated prior to stent placement.According to the complainant, during the procedure, it was difficult to unsheathe the stent.The stent suddenly deployed through the stricture and moved to the proximal part of the colon.A longer wallflex colonic stent was placed stent-in-stent within the initial stent to complete the procedure.There were no patient complications as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good with passing gas.
|
|
Search Alerts/Recalls
|
|
|