• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008K HEMODIALYSIS SYS. OLC/DIASAFE PLUS; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008K HEMODIALYSIS SYS. OLC/DIASAFE PLUS; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 190371
Device Problems Backflow (1064); Mechanical Problem (1384); Reflux within Device (1522)
Patient Problem No Patient Involvement (2645)
Event Date 04/08/2019
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event Description
A biomedical technician (biomed) at a user facility reported that a fresenius 2008k hemodialysis (hd) machine had a saline bag backfill during set-up for use.The backfill was visually observed by staff.The biomed did not report that there were any machine alarms.A patient was not connected to the machine at the time of the incident.The biomed reported that there were no quick disconnects on the drain line and that the drain line length and height were within specification.The biomed believes the saline bag backfill was not due to user error.The biomed confirmed that the unit has received the cbe hardware and software upgrades.The biomed advised he replaced the air separator board and rebuilt valve 24 to resolve the issue and return the machine to service.The machine is back in service without reoccurrence of the reported event.No parts are available for return to the manufacturer.
 
Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius regional equipment specialist (res).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Manufacturer Narrative
Correction: device codes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008K HEMODIALYSIS SYS. OLC/DIASAFE PLUS
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
MDR Report Key8509995
MDR Text Key141789211
Report Number2937457-2019-01097
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
PMA/PMN Number
K994267
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup,Followup
Report Date 04/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number190371
Was Device Available for Evaluation? No
Device Age MO
Initial Date Manufacturer Received 04/08/2019
Initial Date FDA Received04/12/2019
Supplement Dates Manufacturer Received04/23/2019
04/29/2019
Supplement Dates FDA Received04/25/2019
04/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-