Our factory experts evaluated the reported issue within the acceptable risk threshold for the product, however, relevant risk for the site was identified.The syngo.Plaza has been configured to use a primary (nas1) and backup (nas2) network attached storage (nas) mount point for archiving purposes.Due to an unrecoverable hardware failure on the nas1 a restoration of all data was performed from the nas2.However, the nas2 did not contain all archived data as on nas1.8523 series dated within the timeframe 2011-2016 were not in the archive and no longer available.This is considered a data loss of clinical relevant data.The root cause analysis was able to limit down the issue to a manual intervention on the nas2's filesystem, which is not no longer traceable.The syngo.Plaza did not contribute to the missing data in the nas2 archive.There are no injuries attributed to this event.The reported event occured in the (b)(6).
|