• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC HD 1000I SHEARS 36CM SHAFT; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC HD 1000I SHEARS 36CM SHAFT; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HARHD36
Device Problem Loss of Power (1475)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch #r9528y.Investigation summary: the device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched, and with evidence of contact with metal in or out of the operative field.This blade tip portion may have broken off of the device during transport to our analysis site.During functional testing on gen11 an alert screen was displayed.A probable cause of the device stop activating and display an alert screen is blade damage.The device was disassembled to inspect the internal components and no anomalies were found.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" or " removed instrument from patient" followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damage blade can result in a broken blade.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Event Description
While performing a laparoscopic hysterectomy, the hd1000i harmonic stopped working without any reason.It was used normally, with advanced hemostasis button being used very regularly, because the type of structures, but at the vagina/uterus junction approach, the gen11 generator started to present error messages, and hd1000i had to be replaced.Was surgery delayed due to the reported event? yes.If yes, number of minutes: 10 minutes.Action taken when event occurred? harmonic replacement.Was procedure successfully completed? yes.Were fragments generated? no.If yes, were they removed easily without additional intervention? unknown.Patient status/ outcome / consequences: no.Was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown.Is the patient part of a clinical study? unknown.(b)(4).By checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.True.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC HD 1000I SHEARS 36CM SHAFT
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key8515938
MDR Text Key142260942
Report Number3005075853-2019-18125
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10705036015055
UDI-Public10705036015055
Combination Product (y/n)N
PMA/PMN Number
K160752
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 04/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/30/2023
Device Catalogue NumberHARHD36
Device Lot NumberR95538
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2019
Initial Date Manufacturer Received 04/13/2019
Initial Date FDA Received04/15/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/29/2018
Is the Device Single Use? Yes
Type of Device Usage N
Patient Sequence Number1
-
-