Model Number M00546600
Device Problem Material Protrusion/Extrusion (2979)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/27/2019
Event Type  malfunction  
Manufacturer Narrative
Patient's exact age is unknown; however, it was reported that the patient was over the age of 18.(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a spyscope digital access and delivery catheter was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, the working channel sleeve of the spyscope ds protruded at the distal end.Reportedly, a dreamwire and a non bsc ehl probe were inside the spyscope ds when it protruded and there were no reported issues with the accessory devices.The procedure was completed with a second spyscope digital access and delivery catheter.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
Patient's exact age is unknown; however, it was reported that the patient was over the age of 18.A visual assessment was performed.As received, the working channel sleeve (wcs) protruded.Maximum wcs protrusion was observed when the distal tip was articulated by turning the small knob in the counterclockwise direction.The distal tip was cut.The distal cap was removed.The catheter was cut open using the cutting fixture.The wcs was removed.Witness marks were noted on the pebax.The white and clear areas along bond a appeared to show evidence of adhesion.The complaint was consistent with the reported event of working channel sleeve protrusion.Based on investigation results, the root cause of working channel sleeve protrusion is an insufficient bond, particularly the second heat cycle of the working channel sleeve bonding process [bond b].Working channel sleeve protrusion in devices manufactured pre 01mar2018 changes has been determined to be a manufacturing issue, therefore, the complaint investigation conclusion code selected for the wcs protrusion issue is manufacturing deficiency, which indicates that problems were traced to the manufacturing process.An investigation is underway to address this issue.A dhr (device history record) review was performed and did not identify evidence of deviations or non-conformances in the manufacturing processes that could contribute to the complaint.The dhr review confirms that the accepted device met all manufacturing specifications.
 
Event Description
It was reported to boston scientific corporation that a spyscope digital access and delivery catheter was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, the working channel sleeve of the spyscope ds protruded at the distal end.Reportedly, a dreamwire and a non bsc ehl probe were inside the spyscope ds when it protruded and there were no reported issues with the accessory devices.The procedure was completed with a second spyscope digital access and delivery catheter.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPYSCOPE DS
Type of Device
CHOLEDOCHOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8520582
MDR Text Key142200044
Report Number3005099803-2019-01983
Device Sequence Number1
Product Code FBN
UDI-Device Identifier08714729863236
UDI-Public08714729863236
Combination Product (y/n)N
PMA/PMN Number
K142922
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/07/2019
Device Model NumberM00546600
Device Catalogue Number4660
Device Lot Number0021482713
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2019
Initial Date Manufacturer Received 03/27/2019
Initial Date FDA Received04/16/2019
Supplement Dates Manufacturer Received05/15/2019
Supplement Dates FDA Received06/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-