BOSTON SCIENTIFIC CORPORATION SPYSCOPE DS; CHOLEDOCHOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID
|
Back to Search Results |
|
Model Number M00546600 |
Device Problem
Material Protrusion/Extrusion (2979)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/27/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient's exact age is unknown; however, it was reported that the patient was over the age of 18.(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that a spyscope digital access and delivery catheter was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, the working channel sleeve of the spyscope ds protruded at the distal end.Reportedly, a dreamwire and a non bsc ehl probe were inside the spyscope ds when it protruded and there were no reported issues with the accessory devices.The procedure was completed with a second spyscope digital access and delivery catheter.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Manufacturer Narrative
|
Patient's exact age is unknown; however, it was reported that the patient was over the age of 18.A visual assessment was performed.As received, the working channel sleeve (wcs) protruded.Maximum wcs protrusion was observed when the distal tip was articulated by turning the small knob in the counterclockwise direction.The distal tip was cut.The distal cap was removed.The catheter was cut open using the cutting fixture.The wcs was removed.Witness marks were noted on the pebax.The white and clear areas along bond a appeared to show evidence of adhesion.The complaint was consistent with the reported event of working channel sleeve protrusion.Based on investigation results, the root cause of working channel sleeve protrusion is an insufficient bond, particularly the second heat cycle of the working channel sleeve bonding process [bond b].Working channel sleeve protrusion in devices manufactured pre 01mar2018 changes has been determined to be a manufacturing issue, therefore, the complaint investigation conclusion code selected for the wcs protrusion issue is manufacturing deficiency, which indicates that problems were traced to the manufacturing process.An investigation is underway to address this issue.A dhr (device history record) review was performed and did not identify evidence of deviations or non-conformances in the manufacturing processes that could contribute to the complaint.The dhr review confirms that the accepted device met all manufacturing specifications.
|
|
Event Description
|
It was reported to boston scientific corporation that a spyscope digital access and delivery catheter was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, the working channel sleeve of the spyscope ds protruded at the distal end.Reportedly, a dreamwire and a non bsc ehl probe were inside the spyscope ds when it protruded and there were no reported issues with the accessory devices.The procedure was completed with a second spyscope digital access and delivery catheter.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|